TEMP Instruction
TEMP T [20]
Sets the temperature T of the data collection in degrees Celsius. This is
reported to the .CIF file and used to set the default isotropic U values for
all atoms. TEMP must come before all atoms in the .ins file. TEMP also sets
the default X-H bond lengths (see AFIX) which depend slightly on the
temperature because of librational effects. The default C-H bond lengths and
default U-values are rounded to two decimal places so that they may be quoted
more easily.