Re: scatter slit and ADS

Ian Madsen ( ian.madsen@minerals.csiro.au )
Fri, 19 Jun 1998 08:46:04 +1000

At 13:37 18/06/98 +0200, you wrote:
> I am using a difractometer with an automatic divergence slit, with a
>maximum wide of 4 degrees at 2theta=180.
>
> If the scatter slit is 1 degree wide, are the intensities deformed
>when compared with a scatter slit of 4 degrees? The receiving slit was of
>0.5 degrees.
>
> Gustavo Cifredo

The scatter slit should be at least as large as the widest opening of the
ADS - for the situation described above you should be using a 4deg scatter
slit.

There is a good discussion of the issues in :-

Bowden M.E. and Ryan M.J. (1991) Powder Diffraction, 6(2), 78-81.

Cheers

\ | /
___\|/___
(o o)
*-------------oOO--(_)--OOo-----------------*
Ian Madsen
CSIRO Minerals
Box 312
Clayton South 3169
Victoria
AUSTRALIA

Phone +61 3 9545 8785 direct
+61 3 9545 8500 switch
FAX +61 3 9562 8919
Email Ian.Madsen@minerals.csiro.au

XRD laboratory 'tour' starts at
http://www.minerals.csiro.au/public/work/res_equip/tour/begin.htm