Dear Mr, I send you my phase identification result for files SMRR. The interpreted files are in image format HPGL (Hewlett-Packard Graphics Language). However I write you my phase identification: File: SMRR-1 ICDD patterns: 71-880 (Ca5(PO4)3F), 85-335 (SiO2), 86-2348 (Mg(CO3)). In this sample threre are other minor phases still not identified at this stage. File: SMRR-2 ICDD patterns: 88-1173 (TiO2). In this sample I have only identified this minor phase. File: SMRR-3 ICDD patterns: 19-1946 (Thalidomine), 7-710 (6-amino-2thioxo-4(1,3)-pyrimidone hydrate) plus other phase still not identified. File: SMRR-4 ICDD patterns: 5-561 (PbO), 23-333 (Pb5O4SO4) plus minor phases. Software used: DIFFRAC-AT v3.0 from SOCABIM-SIEMENS. ICDD: realase 2000, data sets 1-50 and 70-88 Research experience: Ph D. in Geology (Crystallography). From 1995 up to present working as XRD technicien at the University. I would like to know if we must consider in our interpretation peak position errors due to sample preparation, for instance sample transparency and/or preferred orientation.